GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
7-14 Section 7-4 - Screen Captures and Logs
Error Descriptions:
7-4-3-5-4-2 Software Errors :
Message Description
GFE Ch0 empty-GFE Ch1 empty-CPS_ERROR(FAULT_HVLV). Severity :FatalHW MDC Error.
GFE Ch0 empty-GFE Ch1 empty-CPS_ERROR(FAULT_POWER). Severity :FatalHW MDC Error.
GFE Ch0 empty-GFE Ch1 empty-CPS_ERROR(FAULT_STOP). Severity :FatalHW MDC Error.
GFE Ch0 empty-GFE Ch1 empty-BF_ERROR_A . Severity : FatalHW MCB Error (One of Callisto FPGA)
GFE Ch0 empty-GFE Ch1 empty-BF_ERROR_B . Severity : FatalHW MCB Error (One of Callisto FPGA)
scanctrl ScanControl::startScan: Tx Power Supply is not ready. HV setup timeout after 4127
ms, ScanSequencer NOT started!. Severity : NonFatalSW
MDC Error or MDC Error
Software error
“Error ; EAErrorHandler(2284); Error from DFEManager: DFEManager - RunDfe failed:!,
DFE ERROR: InPort is not connected (2); device "DelayValues_CW.ceDelay". Severity :
NonFatalSW”