Configuration
2026419-033E CIC Pro™ 6-39
Alarm Volume
WARNING
ALARMS — Do NOT rely exclusively on the audible alarm system for Bedside Monitoring. Adjustment of CIC Pro
center alarm volume to a low level or OFF during Bedside Monitoring may result in inability to hear the alarm and
a hazard to the patient. Remember that the most reliable method of Bedside Monitoring combines close personal
surveillance with correct operation of monitoring equipment.
After connecting the monitor to the central station and/or nurse-alert, verify the function of the alarm system.
Repeat this verification periodically, including a check of all connected speakers.
NOTE
All clinical setting values must be received from the GE Clinical Application Specialist (CAS) or from the
nursing director of the care unit. Consult with your CAS or hospital staff about the use of these functions and
if they should be used with your configuration.
Set the alarm volume for the CIC Pro center.
NOTE
The Minimum alarm volume can only be set via the Service mode. You can set the Current alarm volume to
any level above the Minimum via the user or Service mode.
All clinical setting values must be received from the GE Clinical Application Specialist (CAS) or from the
nursing director of the care unit. Consult with your CAS or hospital staff about the use of these functions and
if they should be used with your configuration.
Current: Choose any value from the Minimum alarm volume setting or above. This Current alarm volume
setting determines the actual alarm volume.
Minimum: Choose the minimum alarm volume level. The minimum alarm volume in user mode cannot be
adjusted below this set value.
Real-Time Trend
Graph
Configuration
Enable or disable the display of real-time trend windows in the multi-patient viewer. The real-time trend window
displays the recent patient trends for a maximum of two parameters in a graph format. You can configure the Real-
Time Graph for each patient while in user mode only.
To enable the display of the real-time trend windows in the multi-patient viewer, click the Display Real-Time
Trend Graph checkbox.
Option Function