Configuration
2026419-033E CIC Pro™ 6-43
4. Click the Disable Auto Display Button checkbox in the Auto Display Button
section of the Display Configuration tab.
5. Verify all the Display Configuration settings are the same for the main and
remote CIC Pro centers.
6. Set the Columns and Rows in the Display Configuration tab of the remote CIC
Pro center to match the main CIC Pro center.
7. Click CIC Defaults.
8. From the Mirror Central Display drop-down list, choose the main CIC Pro
center you want mirrored on the remote CIC Pro center.
9. Check the title bar on the remote CIC Pro center to verify that it is mirroring the
correct CIC Pro center.
10. Click Apply to apply your changes without closing the CIC Setup window.
11. Confirm proper operation of the mirror CIC Pro center and the primary CIC Pro
center by completing the appropriate checkout procedures. See Configuration
checkout procedures on page 7-3.