6-74 CIC Pro™ 2026419-033E
Configuration
11. Verify the new settings prior to monitoring patients.
Scenario 2: Add or upgrade a CIC Pro center in a care unit running multiple CIC
Pro centers
When a CIC Pro center is added to or upgraded in a care unit already running other
CIC Pro centers, configuration data common to the care unit is changed in the new or
upgraded CIC Pro center, when it starts up, to match the other CIC Pro centers found
in the care unit network. Configuration data specific only to the new or upgraded CIC
Pro center does not change. Therefore, restore is not useful in this situation. Verify the
new settings prior to monitoring patients.
Scenario 3: Restore multiple CIC Pro centers in a care unit to a previous
configuration
In certain cases, it may be necessary to restore the CIC Pro centers in a care unit to a
previous configuration.
1. Shut down all but one of the CIC Pro centers in the care unit. See Perform safe
shutdown of the CIC Pro center on page 8-31.
This prevents the restore from being overwritten by associated data from other
CIC Pro centers running on the care unit network.
NOTE
The CIC Pro center application must be stopped in order to perform a
restore. Because configuration data is automatically updated to match other
CIC Pro centers found running in the care unit, if other CIC Pro centers are
left running, when the restored CIC Pro center application starts back up,
some configuration data would be overwritten by the automatic update.
2. Stop the CIC Pro center application on the one CIC Pro center that was left
running.
a. From the multi-patient viewer, click Setup CIC.
b. Click the Service Password tab.
c. Type mms_com as the password and press Enter.
d. At the Windows command line prompt, type stop and press Enter.
3. Disconnect the CIC Pro center from the CARESCAPE Network IX and MC
networks.
4. Connect a service PC directly to the CIC Pro center and Log on to the Webmin
service interface remotely via the CARESCAPE network IX network on page 4-
7.
5. Click Configuration > Backup/Restore > Unit Defaults.