8-18 CIC Pro™ 2026419-033E
Troubleshooting
WARNING
LOSS OF MONITORING — Provide alternate patient monitoring
or close observation before performing the system resource reset
procedure. Beds displayed on the CIC Pro center will not be
monitored during the system resource reset.
After system resource reset is complete and the monitoring function
at the CIC Pro center has been restored (about 30 seconds), verify
the correct monitoring state and alarm function.
8. Restart the CIC Pro center. See Perform safe restart of the CIC Pro center on
page 8-30.
Activate licenses (manually) via the Activation Code Summary Sheet
Use this procedure when the license activation codes reside on the printed Activation
Code Summary Sheet.
The printed Activation Code Summary Sheet that matches the serial number of the
CIC Pro center you are the activating the options on is required.
1. If you have not already logged onto Webmin, Log on to the Webmin service
interface on page 4-7.
2. Click Configuration > Licensing.
3. Next to the license you want to activate, type the activation code into the
Activation Code field. See the Activation Code Summary Sheet.
4. Click Activate to activate the license on this CIC Pro center.
NOTE
To remove an activated license, click Remove next to the license you want to
remove.