Equipment overview
2026419-033E CIC Pro™ 2-15
Stores 72 hours of data per patient regardless of license type. Licensing controls
amount of data that is viewable.
Unit Licensing Mode sets which type of license is requested. Supported types:
None (default - 1 hour), 24, 48, 72 hours.
Failure to obtain a license results in no patient data is stored for full disclosure.
Licensing information is stored with data, so restrictions follow data regardless
of where it is viewed, even if viewed in a different care unit with a different Unit
License Mode.
Control
CIC Pro center master and full disclosure
There is one full disclosure master CIC Pro center per care unit. The CIC Pro center
with the lowest MC IP address (126.x.x.x is “lower” than 3.x.x.x) within the care unit
becomes the full disclosure master.
The CIC Pro center master performs the following activities in one minute
increments:
Detects admitted (but not full disclosure data collection) beds within the care
unit, for Auto modes.
Assigns beds (for full disclosure acquisition) to CIC Pro centers within the care
unit.
Applies multi-full disclosure, twin-bed rules, and rover rule.
Identifies CIC Pro centers with full disclosure capability.
NOTE
The CIC Pro center master checks the software versions of the other CIC Pro
centers before trying to identify CIC Pro centers with full disclosure
capability. Software versions earlier than 2.5 are ignored.
Guarantees that the right CIC Pro center is contacted if retrieving data while a
multi-full disclosure condition is occurring and that the multi-full disclosure
condition will be corrected.
NOTE
A CIC Pro center will not act as full disclosure master during the first minute
after it starts up. Instead, the latest data is first collected from the other CIC
Pro centers on the network, and this data determines which CIC Pro center
becomes master. Multiple masters on a network might occur, but only briefly
(this will self-correct within a couple of minutes).
Minute rule or the Offline Storage setting
For more information, see Set the full disclosure defaults on page 6-47.
The Offline Storage setting determines the length of time the CIC Pro center will
maintain full disclosure data for a bed from which it has stopped receiving data (e.g.,
a NO COMM condition).
If the NO COMM condition ends within the Offline Storage setting time frame,
full disclosure data collection for the bed continues.