2-16 CIC Pro™ 2026419-033E
Equipment overview
If a NO COMM condition exceeds the Offline Storage setting time frame, all
data for the bed is deleted. If (while in Auto mode) the bed comes back online
(the NO COMM condition ends) after this point, the bed is reassigned as a new
and different bed and new data collection is started for the bed.
When a CIC Pro center starts up, it determines when full disclosure data was last
received from the assigned beds.
If the latest data for a bed:
Is not older than the Offline Storage setting, then full disclosure data
collection continues for the bed.
Is older than the Offline Storage setting, existing data associated with the
bed is deleted, and new data collection is started for the bed.
Multi-full disclosure rule
If multiple CIC Pro centers are running in a care unit, the multi-full disclosure rule
allows the switching of full disclosure data collection from one CIC Pro center to
another, if a CIC Pro center goes offline (e.g., reboot, shutdown, etc.).
The following examples describe how the multi-full disclosure rule works when using
two CIC Pro centers (CICA and CICB).
NOTE
In the examples, start up mode is Auto, and only one bed (BED1) is used.
Example 1:
1. CICA goes offline and stops full disclosure data collection on BED1. This causes
CICB to begin full disclosure data collection on BED1.
NOTE
The master CIC Pro center detects CICA going offline, and switches full
disclosure data collection over to CICB, within a minute of the offline event
occurring.
2. CICA comes back online within the Offline Storage setting time frame, and
continues full disclosure data collection on BED1 again. At this point, CICA and
CICB are both collecting full disclosure data on BED1.
3. The multi-full disclosure rule does not allow more than one CIC Pro center to
collect full disclosure data from the same bed at the same time--only the data that
goes the farthest back in time (CICA) is kept. Because of this, CICA continues
collecting full disclosure data on BED1, and CICB stops.