GENERAL
Parameters valid for all selected cycles.
STANDARD
VALUE (type)
TI15 G DRAIN TO T1T2: value 0-2 min
General drain to tank 1 or tank 2. (draining time used to drain the washer-
disinfector to one of the tanks without filling and then saving the rest in the
sump.
Used with drainage neutralizer and descaling).
Starts directly when the drain pump starts.
45 sec (P)
TI09 G DR SHORT TIME: value 0-2 min
Short time set point for general drainage. (draining timer used when the
washer already is partly drained before e.g. after neutralization of remaining
content when using drain neutralization.)
Starts directly when the drain pump starts if CHAMBER LEVEL H has lost
its signal.
20 sec (P)
PF29 G DRAIN NEUTR: value 0.1
General drain neutralization (Parameter for selecting or deselecting the
drain neutralization.)
Note: Check that the correct chemicals are used.
0 (P)
LI76 G A NEUTR pH L: value 6-10
General alarm, neutralizing pH low. The limit is used when using drain
neutralization with pH monitoring. If the pH value is below the limit after
drain neutralization, an alarm is generated and the process fails.
7 (P)
LI77 G A NEUTR pH H: value 7-11
General alarm, neutralizing pH high. The limit is used when using drain
neutralization with pH monitoring. If the pH value is over the limit after drain
neutralization, an alarm is generated and the process fails.
9 (P)
TI08 G LOWPOINT DRAIN: value 0-10 min
General low point for drainage. Used when the washer is equipped with an
extra low point drain valve in order to drain the washer completely.
2 min (P)
LI41 G TEMP DRAIN L: value 0-70°C
General temperature for drain low-low. When the drainage cooling is
installed, this is the temperature when the cooling valves are shut off when
falling below the limit.
30°C (P)
LI42 G TEMP DRAIN H: value 30-70°C
General temperature for drain high. When the drainage cooling system is
installed, this is the temperature at which the cooling valve opens when the
limit is exceeded and the drain pump restarts when falling below the limit.
40°C (P)
LI43 G TEMP DRAIN HH: value 30-70°C
General temperature for drain high-high. When the drainage cooling is
installed, this is the temperature at which the drain pump stops because
the limit is exceeded.
50°C (P)
CR06 G TEMP BOOST SP: value 70-93°C
General temperature for booster tank. This temperature is maintained in
the booster tank during standby unless the time-out booster TI139 has
expired.
90°C (P)
Page 79 of 166
<Doc_TEC><Doc_6001341502><Rel.A><Lang_en>

Table of Contents

Other manuals for Getinge 9125

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge 9125 and is the answer not in the manual?

Getinge 9125 Specifications

General IconGeneral
BrandGetinge
Model9125
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals