GENERAL
Parameters valid for all selected cycles.
STANDARD
VALUE (type)
CT12 G TANK1 FILL: value 5-100
Filling for general tank 1. This is the number of coils in which the water can
be recycled before the system is automatically drained and restarted with
clean water.
10 (P)
CT13 G TANK2 FILL: value 5-100
Filling for general tank 2. This is the number of coils in which the water can
be recycled before the system is automatically drained and restarted with
clean water.
10 (P)
TI39 G TIME OUT BOOSTER: value 0-99 h
General time-out for booster tank. This is the time that passes before the
booster tank switches to standby depending on temperature and level
control.
10 h (P)
TI50 G CH LEVEL H DEL: value 5 sec -1 min
General level for high delay. Delay time during water filling to get a stable
water level during operation.
5 sec (P)
PF36 G PASS THROUGH: value 0.1
General pass through. Parameter for activating and deactivating pass
through. Its only purpose is to create a pass through cycle.
0 (P)
PF40 G START PASSWORD: value 0.1
General start password. Parameter for activating or deactivating start
password when starting the cycle.
0 (P)
Page 80 of 166
<Doc_TEC><Doc_6001341502><Rel.A><Lang_en>

Table of Contents

Other manuals for Getinge 9125

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge 9125 and is the answer not in the manual?

Getinge 9125 Specifications

General IconGeneral
BrandGetinge
Model9125
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals