WARNING! To avoid the risk of electric
shock, this equipment must only be
connected to a supply mains with protective
earth.
CAUTION: Federal law restricts this device
to sale by or on the order of a physician.
2.2 Connection
A full System checkout procedure must be
performed at least once a day.
The system must never be left unattended
while connected to a patient.
Electronic accessories and auxiliary
equipment other than the vaporizers must
not be connected or disconnected during
operation or when the system is plugged
into a mains power outlet. Such connection
or disconnection may interfere with the
functioning of the system.
Supplied gases shall meet the requirements
for medical grade gases according to
applicable standards.
The backup gas supply should only be
turned ON (valves open) when the backup
gas supply is in use, or during System
checkout.
FLOW-i 4.2, User's Manual