Patient cases should not be started if only
Leakage or Vaporizer checks have been
performed since system startup or during the
last 24 hours period.
Press the 'Results' touch pad to display a
window describing the system checkout
results in more detail.
4.5 Understanding results
01-01 13 00
32
1
In Standby mode, the central area of the
screen displays the result of performed system
checks, along with date, time and outcome.
The symbol (1) can be either green, yellow or
red, depending on the current status of system
checkouts. A detailed view is obtained by
pressing 'View results' (2). A full system
checkout, leakage check or vaporizer check
is initiated by pressing the corresponding
touch pad (3).
Symbol color codes correspond to the
following:
Red symbolYellow symbolGreen symbol
Failed. One or several checks
failed during the System
checkout/Leakage
check/Vaporizer check. Make
appropriate adjustments and redo
check.
More than 24 hours since a
successful full system checkout
was performed. Perform a System
checkout.
Not performed. A patient case
can be started, but this is not
recommended. Perform a System
checkout.
Passed. A patient case can be
started.
FLOW-i 4.2, User's Manual
61
| Startup and system checkout |
4 |