EasyManuals Logo
Home>Getinge>Medical Equipment>Maquet FLOW-i 4.2

Getinge Maquet FLOW-i 4.2 User Manual

Getinge Maquet FLOW-i 4.2
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #146 background imageLoading...
Page #146 background image
6.15.4 Warm-up
During startup, the gas analyzer passes
through an initialization and warm-up period,
where the etCO2, O2, and agent values are
displayed as '***' for 60 seconds before ISO
standard accuracy is achieved and the unit is
operational. Delivery of anesthetics and N2O
can begin once the analyzer is operational.
Normal/full accuracy is reached within 10
minutes after power on.
6.16 Lift (C30 model only)
With the mains power connected, the lift
function is operated by means of the Up and
Down buttons located to the right of the
ON/OFF button:
The lift has a 10 % duty cycle; it may be used
continuously for a 2 minute period, after which
it must cool down for 18 minutes. If the lift is
used before it has cooled down completely,
the time of continuous use will be less than 2
minutes.
The maximum capacity is 150 kg (330 lbs),
equivalent to approx. 30 kg (66 lbs) of auxiliary
equipment.
For a detailed description of equipment weight
restrictions refer to Chapter 15, page 268.
146
FLOW-i 4.2, User's Manual
| 6
| Breathing system |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals