EasyManuals Logo

Getinge Maquet FLOW-i 4.2 User Manual

Getinge Maquet FLOW-i 4.2
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #253 background imageLoading...
Page #253 background image
12.9 Drying alternatives
Dismantle
Wipe off/discard
Rinse
DryDisinfectant
Rinse
Washer-
disinfector
Dry
Assemble
System checkout
Steam
autoclave
The patient cassette, volume reflector and
associated parts, must be dried before use.
If the assembled system parts are not dry, the
System checkout may fail.
Drying may not be necessary, or the drying
time may be reduced, if the washer-disinfector
has a drying phase.
Prior to any drying procedure, manually
remove as much water as possible by shaking
and tilting the parts.
FLOW-i 4.2, User's Manual
253
| Routine cleaning and maintenance |
12 |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals