12.15 Starting MCare remote services
There are several methods to start the remote
service process:
A remote service touch pad is available on
the log-in screen when selecting
Menu/Service & Settings.
The startup configuration can be set so that
the system connects to the MCare portal
at specified occasions:
OFF-
- Prompts start of remote services after a
completed SCO.
- Always performs remote services after a
completed SCO.
- Always performs remote services after a
completed patient case.
12.15.1 Startup configuration
The startup configuration settings for remote
service is available in Service & Settings, see
Chapter 9, page 174.
12.15.2 Running MCare remote services
When remote services is active and running,
a status window is displayed on the control
panel. It shows the progress of transfer, as
well as any issues with the process. The
window closes automatically upon completion.
To start a patient case when a remote services
session is active, press ‘Cancel’ in the remote
services status window. A patient case cannot
be started when remote services is active and
running.
FLOW-i 4.2, User's Manual