4.3 Leakage check
Performs a Leakage check of the manual and
automatic breathing circuits. It shall be
performed after each change of tubings,
breathing bags or filters.
01-01 13 00
2 4
1 3
Leakage check preparatory phase contains
four steps, equivalent to steps 1, 2, 3 and 6
performed during the Full system check, see
section 4.2.1.
The automatic check sequence includes
automatic and manual breathing circuit
leakage checks.
01-01 13 00
1 2 3
A summary of the results is displayed after the
automatic check. Press the Standby touch
pad to enter Standby mode or press any of
the touch pads marked 1-3 to repeat any of
the checks.
The circuit compliance compensation is
re-calculated during the Leakage check and
the log is updated with the new measured
values for leakage.
FLOW-i 4.2, User's Manual
59
| Startup and system checkout |
4 |