6.8 Patient tubing
Maquet recommends using patient tubings
measuring from 1.2 to 3 meters in length.
It is recommended only to use original tubes
from MAQUET. Soft tubing or tubing with high
internal resistance may negatively effect the
performance of the system.
A new Leakage check must be performed after
every change of tubings, bags or filters, e.g.
changing from adult to infant tubings. This to
prevent leakage and update the circuit
compliance compensation.
6.9 Manual breathing bag
A
The manual breathing bag is connected to the
marked port (A) on the patient cassette.
The manual breathing bag should comply with
ISO 5362:2006.
124
FLOW-i 4.2, User's Manual