14.2 Electromagnetic emissions
Guidance and manufacturer's declaration - electromagnetic emissions
The FLOW-i is intended for use in the electromagnetic environment specified below. The customer or
the user of the FLOW-i should assure that it is used in such an environment.
Electromagnetic environment - guidanceComplianceEmissions test
The FLOW-i uses RF energy only for its internal function.
Therefore its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Group 1RF-emissions
CISPR11
The FLOW-i is suitable to use in all establishments other than
domestic and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
Class ARF-emissions
CISPR11
Not applicableHarmonic
emissions
IEC 61000-3-2
Not applicableVoltage
fluctuations/flicker
emissions
IEC 61000-3-3
FLOW-i 4.2, User's Manual