6.17.1 Docking and undocking
The C40 system must always stand on the
floor when it is being docked or undocked
onto or from the ceiling pendant.
The C40 system is intended for transport
inside the operation room only when undocked
from the ceiling pendant. Transportation to
other locations requires the use of a suitable
cart or trolley according to hospital routines.
All personnel working with the system shall be
well acquainted with the docking/undocking
procedure so that, in the event of an
emergency, the system can be undocked and
removed.
WARNING!
Ensure that the system is properly docked
and secured before manoeuvring the C40
using the ceiling pendant.
This device, when removed from its wall or
ceiling mount, does not meet the stability
requirements of ISO 80601-2-13 and
IEC 60601-1 respectively. Special caution
has to be taken.
CAUTION: Disconnect all electrical cables
and gas connections before undocking the
system from the ceiling pendant.
148
FLOW-i 4.2, User's Manual
| 6
| Breathing system |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals