14.4 Recommended separation
distances
Recommended separation distances between portable and mobile RF communications equipment
and the FLOW-i.
The FLOW-i is intended for use in an electromagnetic environment in which radiated RF-disturbances
are controlled. The customer or the user of the FLOW-i can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the FLOW-i as recommended below, according to the maximum power of the
communications equipment.
Separation distance according to frequency of transmitter (m)Rated maximum
output power of
transmitter (W)
800 MHz to
2.5 GHz
d = 2.3
P
80 MHz to 800 MHz
d = 1.2
P
150 kHz to 80 MHz
in ISM bands.
d = 1.2
P
150 kHz to 80 MHz
outside ISM bands.
d = 0.35 P
0.230.120.120.040.01
0.730.380.380.110.1
2.301.201.200.351
7.273.793.791.1110
23.0012.0012.003.50100
See the notes in the previous tables.
298
FLOW-i 4.2, User's Manual
| 14
| EMC Declaration |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals