14.3 Electromagnetic immunity
Guidance and Manufacturer´s Declaration - electromagnetic immunity
The FLOW-i is intended for use in the electromagnetic environment specified below. The customer or
the user of the FLOW-i should assure that it is used in such an environment.
Electromagnetic
environment - guidance
FLOW-i Compliance
level
IEC 60601 test levelImmunity test
Floors should be made
of concrete, wood or
ceramic tile. If floors are
covered with synthetic
material, then the relative
humidity should be at
least 30%.
± 6 kV Contact discharge
± 8 kV Air discharge
± 6 kV Contact discharge
± 8 kV Air discharge
Electrostatic discharge
(ESD)
IEC 61000-4-2
Mains power quality
should be that of a
typical hospital
environment.
+
2 kV power supply
lines
+
1 kV input/output lines
+
2 kV power supply
lines
+
1 kV input/output lines
Electrical fast
transient/burst
IEC 61000-4-4
Mains power quality
should be that of a
typical hospital
environment.
± 1 kV line to line
± 2 kV line(s) to earth
± 1 kV line to line
± 2 kV line(s) to earth
Surge IEC 61000-4-5
The power frequency of
magnetic fields should
be at levels
characteristic of a typical
location in a typical
hospital environment.
30 A/m3 A/mPower frequency
(50/60 Hz) magnetic
fields.
IEC 61000-4-8
294
FLOW-i 4.2, User's Manual
| 14
| EMC Declaration |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals