Mains power quality
should be that of a
typical hospital
environment. The
anesthesia system
requires continuous
operation during power
mains interruptions.
FLOW-i is equipped with
internal batteries. The
responsible
organisation/user shall
make sure that batteries
are replaced before their
lifetime has expired,
according to the
instructions of use.
>95% reduction for
0.5 cycle, 60% reduction
for 5 cycles, 30%
reduction for 25 cycles
and >95% reduction for
5 s.
>95% reduction for
0.5 cycle, 60% reduction
for 5 cycles, 30%
reduction for 25 cycles
and >95% reduction for
5 s.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines.
IEC 61000-4-11
FLOW-i 4.2, User's Manual
295
| EMC Declaration |
14 |