5. Remove the cleaning adaptor, carefully
shake and tilt the cassette, turn upside
down and repeat.
12.6.2 Volume reflector
Connect running water to one inlet of the
volume reflector and let water flow through for
at least 1 minute. Make sure no air pockets
remain in the internal compartment. Place the
volume reflector standing on its long end
according to the illustration.
12.6.3 Volume reflector adaptor and safety
valve
Rinse the volume reflector adaptor and safety
valve (if removed) under running water for at
least one minute.
12.7 Disinfection procedures
Dismantle
Wipe off/discard
Rinse
DryDisinfectant
Rinse
Washer-
disinfector
Dry
Assemble
System checkout
Steam
autoclave
This section describes two independent
methods of disinfecting the patient cassette,
volume reflector and associated parts:
Washer-disinfector
Treating the parts with disinfectant fluid
The patient cassette is a precision instrument
and must be handled carefully.
248
FLOW-i 4.2, User's Manual
| 12
| Routine cleaning and maintenance |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals