12.6.1 Patient cassette
Use the required cleaning adaptor when
rinsing the patient cassette:
1. Connect the cleaning adapter to fit into
the back of the patient cassette.
2. Connect the cleaning adapter to a source
of running water (flow > 10 l/min) using
connective tubing. Let water flow through
the cassette for at least one minute.
3. Place the patient cassette upside down
and again let water flow through the
cassette for at least one minute.
4. Attach the connective tubing to the
expiratory connection and let water flow
through the cassette for at least one
minute. Repeat for the inspiratory
connection.
FLOW-i 4.2, User's Manual
247
| Routine cleaning and maintenance |
12 |