3.5 Vaporizer unit
The vaporizer unit holds one to two vaporizers,
where vaporizers can be selected for the
following agents:
Isoflurane
Sevoflurane
Desflurane
3.6 Emergency ventilation system
30
60
80
APL
mbar / cmH
2
O
O
2
I/min
O
2
I/min
2
4
6
8
10
SP
Emergency ventilation
In case of a total power (i.e. mains power and
battery) or system failure, this system will allow
the patient to be manually ventilated. The
emergency ventilation system comprises:
1. Activation switch
2. O
2
gas supply and flowmeter, graded up
to 10 l/min.
3. Mechanical APL
APL
mbar / cmH
2
O
O
2
I/min
O
2
I/min
2
4
6
8
10
Not for use with AFGO
1 3
2
WARNING! If the emergency ventilation
system is activated while the anesthesia
system is in operation, the anesthesia
system will be shutdown.
26
FLOW-i 4.2, User's Manual