13.17 Gas analyzer
Gas analyzer
Paramagnetic sensorO
2
Measuring technology
IR sensorAgents, CO
2
, N
2
O
Within 60 sISO standard accuracyWarm-up time
Within 10 minutesFull accuracy
225 ml/min +/- 10% (Return to circuit), BTPS conditionSampling flow and tolerance
• Length: 2.0 m
• Inner diameter: 0.05"
Sampling line
Measured parameters
2 - 100 breaths/minuteResp. rate
± 1 breath/minute<60 breaths/minute
Unspecified>60 breaths/minute
Respiration rate measurement
accuracy
Gas measurement accuracy
Interference [%
ABS
]Accuracy [%
ABS
]:Gas conc.
± 10 - 25% 0.1N
2
O
± 225 - 80% 0CO
2
± 380 - 100% 0.3Any agent
O
2
Interference [%
ABS
]Accuracy [%
ABS
]:Gas conc.
± 20 - 20% 0CO
2
± 320 - 80% 0O
2
0Any agent
N
2
O
Interference [%
ABS
]Accuracy [%
ABS
]:Gas conc.
± 0.10 - 1% 0.1N
2
O
± 0.21 - 5% 0.1O
2
± 0.35 - 7% 0.3Any agent
± 0.57 - 10%
Unspecified> 10%
CO
2
Interference [%
ABS
]Accuracy [%
ABS
]:Gas conc.
± 0.150 - 1% 0CO
2
± 0.21 - 5% 0.1N
2
O
Unspecified> 5% 0.1O
2
0.22nd agent
Isoflurane
Interference [%
ABS
]Accuracy [%
ABS
]:Gas conc.
± 0.150 - 1% 0CO
2
± 0.21 - 5% 0.1N
2
O
± 0.45 - 8% 0.1O
2
Unspecified> 8% 0.22nd agent
Sevoflurane
280
FLOW-i 4.2, User's Manual
| 13
| Technical specifications |