13.16 Vaporizer
Vaporizer
Isoflurane, Sevoflurane and DesfluraneAgents
Electronic InjectorType
ca. 7.1 lbsWeight (full)
2.8" x 8.5" x 7.0"Dimensions
300 mlAgent capacity
30 ml (triggering the low level alarm)Residual capacity
0, 0.3-5%, OFFIsofluraneSetting range
0, 0.3-8%, OFFSevoflurane
0, 1.0-18%, OFFDesflurane
±15% of set value or ±5% of maximum possible user setting
(whichever is greater)
Accuracy
Maquet filling adapterIsofluraneFilling system
Quik Fil
®
and Maquet filling
adapter
Sevoflurane
SAFE-FIL
TM
Desflurane
Approx. 4 ml/sVaporizer filling speed
Overfilling prevention systems built into the vaporizerOverfill protection
Optical and electronicTank liquid level