12.6 Rinsing before disinfection
Dismantle
Wipe off/discard
Rinse
DryDisinfectant
Rinse
Washer-
disinfector
Dry
Assemble
System checkout
Steam
autoclave
Rinse before disinfection as soon as possible
after use to improve the disinfection and
reduce the risk for cross-contamination
between patients. Begin by rinsing the exterior
surfaces under running water. Use the
cleaning adaptors as described below to rinse
the interior compartments of the equipment.
Rinse the parts thoroughly in water
(<35°C/95°F). Let the water flow through the
parts to remove organic matter, e.g. blood,
and other residues. Rinsing with water warmer
than 35°C (95°F) may cause organic material
to coagulate.
If the step of rinsing before disinfecting is not
included in the washer disinfection program,
or if the washer disinfector will not be used
immediately, then rinse the equipment
manually.
Visually check the cleaning adapters for cracks
and fissures. Discard broken adapters.
246
FLOW-i 4.2, User's Manual
| 12
| Routine cleaning and maintenance |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals