3. To select a vaporizer, press the required
vaporizer type and confirm by pressing
Accept. This activates the vaporizer. The
agent name is shown on the Vaporizer
setting touch pad.
If the new vaporizer contains another agent,
the suggested agent concentration displayed
on the control panel is based on the equivalent
MAC value of the previous agent.
When a second vaporizer is activated, the
other vaporizer is automatically de-activated;
the agent concentration is set to OFF, the
vaporizer is depressurized, flushed and
unlocked.
It is possible to manually set the agent
concentration to OFF, and the vaporizer to
standby, using the Agent concentration touch
pad, see page 141 for details.
5.4 Ventilation settings
The system can be run using manual (MAN)
or automatic (AUTO) ventilation. Ventilation
modes available for automatic ventilation
include the following:
Pressure Control (PC)
Volume Control (VC)
Pressure Support (PS)
Manual ventilation modes include manual
ventilation using the circle system and manual
ventilation using an external breathing system
connected to the AFGO outlet (option).
The system also allows for the patient category
(Infant or Adult) to be specified, affecting
parameter setting limits and step values.
FLOW-i 4.2, User's Manual
83
| System functionality |
5 |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals