4.6 Standby mode
01-01 13 00
The Standby screen provides a summary of
the System status, the current gas/ventilation
settings and the selected ventilation mode. If
a vaporizer is connected and selected, the
name of the agent is presented on the screen.
Listed below are examples of functions that
can be performed in Standby mode either by
pressing the active touch pads, or by
accessing the menu using the Menu button.
The gas and ventilation settings can be
preset before starting a patient case.
Define patient settings.
Any or all system checks can be started;
System checkout, Leakage test and
Vaporizer test.
The Service and settings menu can be
accessed.
A patient case can be started.
Alarm limits can be preset before starting
a patient case.
Test results of the latest performed checks
can be reviewed by pressing the results
touch pad.
If the System checkout was bypassed, or if
the System checkout failed, a corresponding
system message will be displayed in the
message area.
When a patient case is started using AUTO
ventilation or a switch is made during Run
mode to AUTO ventilation, the preset gas and
ventilation settings are activated.
62
FLOW-i 4.2, User's Manual
| 4
| Startup and system checkout |