13.8 Fresh gas flow
Fresh gas flow
Gas mix:
Electronic Servo controlledAir/O
2
Electronic Servo controlledO
2
/N
2
O
MAN = 0.3 - 20 l/min
AUTO = 0.3 - 20 l/min (FGF delivery depending on set MV)
AFGO = 1.0 - 20 l/min
Fresh gas flow range
SelectableFresh gas O
2
/air Flow
(numerical/bargraph)
SelectableFresh gas O
2
/N
2
O Flow
(numerical/bargraph)
21% - 100% ±5%
Air/O
2
Preset O
2
concentration
28% – 100% ±5%
O
2
/N
2
O
1%Setting resolution
Approximately 55 l/min
2 cmH
2
O expiratory resistance when APL is set to SP
O
2
Flush
13.9 Anesthesia gas scavenging (AGS)
Anesthesia gas scavenging (AGS)
Passive system (including a flow indicator) integrated into the systemType
Minimum 25 l/min (STPD), or 10 l/min (STPD) over the set minute
volume, whichever is greater.
Scavenging flow
30 mm ISO taper
DISS EVAC
12.7 mm / ½" in hose Barb
25 mm / 1" Barb
AGA EVAC
WAGD-to-Vacuum connector
22 mm out. diam. connector and 22 mm int. diam. connection
tube
Outlet connection type options
FLOW-i 4.2, User's Manual
273
| Technical specifications |
13 |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals