To remove water from the volume reflector the
following steps are recommended:
1. Hold the volume reflector horizontally and
tilt the device so that water flows from the
two openings into the sink.
2. Hold the volume reflector vertically, facing
the top side (smooth side). Slowly rotate
clockwise at least four times until the flow
of water is minimal. Beware of residual
hot water if the equipment has been
cleaned using a washer-disinfector.
1
2
3
4
3. Visually check for residual water in the
channels by holding the volume reflector
in front of a light source.
254
FLOW-i 4.2, User's Manual
| 12
| Routine cleaning and maintenance |