The 'Insp Exp hold' dialog can be accessed
by either of the following:
1. Press the 'Calculate static measurements'
touch pad, located on page three in the
ventilation measurements area.
2. Press anywhere on page three in the
ventilation measurements area.
3. Press the 'Menu' membrane button and
select the 'Insp hold / exp hold' from the
menu.
312
01-01 13 00
3
PEEPtot
16:29
PEEP
(cmH
2
O)
C dyn
(cmH
2
O)
Tc
16:29
E
16:29
C static
16:29
3
67
15.0
0.15
314
1
2
3
The 'Insp. Exp. Hold' dialog has two active
touch pads:
1
2
3
1. Insp. hold
2. Exp. hold
3. Suggested holding time to calculate static
measurements.
Press the appropriate touch pad to initiate the
desired function. The function is active as long
as the touch pad remains pressed down.
Insp./exp hold can also be initiated using the
rotary knob. Select the desired function and
press down the rotary knob. The function is
active as long as the rotary knob remains
pressed down.