Storage
Containers holding unused absorbers should
be stored in a clean, dry environment, and at
temperatures between 0°C (32
o
F) and 35°C
(95
o
C). Storage at higher temperatures can
result in reduced efficiency and service life
due to moisture loss. When correctly stored,
canisters will maintain absorption capacity for
a period of two years.
Canisters must NOT be stored where they can
become subject to the following:
Direct strong sunlight.
Close proximity to incompatible chemicals
or acids.
Partial or total water immersion.
Environment with abnormal concentrations
of carbon dioxide, hydrogen sulphide or
other acidic gases.
Excessive stacking loads - maximum two
pallets high.
128
FLOW-i 4.2, User's Manual
| 6
| Breathing system |