EasyManuals Logo
Home>Getinge>Medical Equipment>Maquet FLOW-i 4.2

Getinge Maquet FLOW-i 4.2 User Manual

Getinge Maquet FLOW-i 4.2
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
The use of O
2
and Air gas outlets may,
depending on central gas supply pressure
and ventilation settings, affect ventilation
performance.
Reprocessing parts labelled 'Single-use'
will degrade biocompatibility and
cleanliness.
WARNINGS!
The use of other accessories,
transducers and cables other than those
specified by MAQUET may result in
increased emissions or decreased
immunity (EMC) of the system.
No other electrical equipment other than
those described in this user's manual
may be placed on, or in the immediate
vicinity of the system.
The patient monitor power connector
must only be used for patient monitors
or equipment mounted on the top shelf.
CAUTIONS:
Only MAQUET recommended
accessories, supplies and auxiliary
equipment must be connected to or used
in conjunction with the system. Use of
other unvalidated accessories, supplies
and auxiliary equipment may impair the
performance and safety of the system.
To prevent the system from tilting, follow
the restrictions for patient monitors and
auxiliary equipment specified in Chapter
15, page 268. Maximum torque on the
side rails by the table is 20 Nm.
Equipment placed on shafts or rails must
not restrict opening of the Emergency
Ventilation cover.
FLOW-i 4.2, User's Manual
15
| Important |
2 |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals