8.10.2 System checkout
Menu - System checkout
Result
Brings up a detailed summary of the last System
check, including leakage info and circuit compliance
calculations, and the result of the System check
before that.
01-01 13 00
Full check
Starts a full system checkout.
Full check, Leakage check and Vaporizer check are
only available in Standby mode
See Chapter 4 for details for details on all system
checks.
01-01 13 00
Leakage check
Performs a leakage check of the breathing system.
The circuit compliance compensation is
re-calculated during the Leakage check and the log
is updated with the new measured values for
leakage.
166
FLOW-i 4.2, User's Manual
| 8
| Membrane buttons |