12.4 Preparations and dismantling
Dismantle
Wipe off/discard
Rinse
DryDisinfectant
Rinse
Washer-
disinfector
Dry
Assemble
System checkout
Steam
autoclave
12.4.1 Remove external appliances
1. Shutdown the system.
2. Disconnect the system from the power
and gas supply.
3. Disconnect any optional equipment from
the anesthesia system.
4. Remove the water trap and gas sampling
line. Discard/empty any water present in
the water trap.
5. Remove patient tubing, Y-piece, and
manual breathing bag.
238
FLOW-i 4.2, User's Manual
| 12
| Routine cleaning and maintenance |