Technical alarms
Contact Maquet serviceControl panel disconnectedTE 81
Contact Maquet serviceControl panel high temperatureTE 606
System is still usable. Alarm
sound signals will be different,
non-alarm sound signals will not
be generated.
Audible alarm errorTE 604
Check video cable connection.
Contact Maquet service.
Communication errorTE 607, 611
Check Control panel.
Contact Maquet service.
Key press timeoutTE 603
Alarms and patient safety
Contact Maquet serviceRe-specify alarm limitsTE 11
Contact Maquet serviceEmergency ventilation On/Off
error
TE 62
Check alarm profileAlarm limit error, check alarm
profile
TE 30
Switch to emergency ventilationAPL valve malfunctionTE 57
Replace cylinderBackup cylinder pressure
unknown: Air/O
2
/N
2
O
TE 85-87
Miscellaneous
Contact Maquet serviceFan malfunctionTE 17
Contact Maquet serviceSystem temperature: HighTE 78
Contact Maquet serviceSettings lost. Restart the system!---
288
FLOW-i 4.2, User's Manual
| 13
| Technical specifications |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals