Positioning the light
56 / 106
VOLISTA
IFU 01781 EN 19
4.3 Positioning the light
4.3.1 Fitting the sterilisable handle
WARNING!
Risk of infection
If the sterile handle is not in good condition, there is a risk that particles
could fall from it into the sterile environment.
After each sterilisation and before using a sterilisable handle again, check
that there are no cracks.
WARNING!
Risk of infection
The sterilisable handles are the only parts of the device that can be sterilised.
Any contact by the sterile team with another surface results in a risk of infec-
tion. Any contact by non-sterile personnel with these handles results in a risk
of infection.
During the procedure, the sterile team must handle the device using the ster-
ilisable handles. On an HLX handle, the locking button is not sterile. Non-
sterile personnel must not come into contact with the sterilisable handles.
4.3.1.1 Installing and removing the STG PSX sterilisable handle
Fig.55: Installing the STGPSX sterilisable handle
Installing a sterilisable handle on the light-
head
1. Inspect the handle and check for cracks or
soiling.
2. Fit the handle to the mount.
Ø A click is heard.
3. Rotate the handle until its rotation is
locked.
Ø The handle is now locked in place and
ready for use.