Troubleshooting
Potential failures and troubleshooting
5
VOLISTA
IFU 01781 EN 19
87 / 106
5.2 Potential failures and troubleshooting
Mechanical components
Problem Likely cause Corrective action
The sterilisable handle does
not click into place correctly
Sterilisation parameters (tem-
perature, time) exceeded
Check the operation of the
handle and in particular the
locking mechanism (audible
click).
The maximum handle service
life has expired or the handle
is twisted or bent.
Replace the handle
The lighthead drifts Suspension tube not vertical Check tube verticality and ceil-
ing structure
Load-bearing structure un-
stable
Contact the Getinge technical
department
Locking screw incorrectly ad-
justed.
Have the locking screw adjus-
ted by a trained technician
Lighthead moves too easily or
is difficult to move.
Locking screw incorrectly ad-
justed.
Have the locking screw adjus-
ted by a trained technician
Insufficient lubrication Contact the Getinge technical
department
Tab.23: Mechanical anomalies and malfunctions
Light system
Problem Likely cause Corrective action
The lighthead does not turn
on.
Power cut Check if other devices on the
same supply network are oper-
ating
Does not switch over to
backup
Contact the Getinge technical
department
Other reason Contact the Getinge technical
department
The lighthead does not turn
off.
Faulty connection between
power supply and lighthead
Contact the Getinge technical
department
None of the lightheads light up Each lighthead has its own
control
Check the LED on the Lexan
panel of each lighthead
A group of LEDs or one LED
does not come on
The LED board or the LED is
defective
Contact the Getinge technical
department
The electronic circuit board
does not communicate with
the LED board
Contact the Getinge technical
department
The light flickers. Incorrectly installed Contact the Getinge technical
department
Tab.24: Optical anomalies and malfunctions