Troubleshooting
Potential failures and troubleshooting
88 / 106
VOLISTA
IFU 01781 EN 19
Problem Likely cause Corrective action
Ambient light does not turn on The button is defective Contact the Getinge technical
department
Faulty connection between
power supply and lighthead
Contact the Getinge technical
department
AIM mode does not turn on This function is not available
on the lighthead
Check whether “AIM” is
marked on the label
The button is defective Contact the Getinge technical
department
Tab.24: Optical anomalies and malfunctions
Other
Problem Likely cause Corrective action
The two lightheads are con-
trolled simultaneously in error
Faulty connection between
power supply and lighthead
Contact the Getinge technical
department
No image after starting the
OHDII AIR03 QL FHD camera
Communication problem
1. Check on the touchscreen
control panel whether the
camera is on
2. Disconnect and reconnect
the receiver power cable
3. Re-register the camera
4. Contact the Getinge tech-
nical department
Image lost for more than
20seconds
Interference with another sys-
tem
1. The system should reset it-
self within around
20seconds
2. Modify the zoom level
3. Contact the Getinge tech-
nical department
Tab.25: Other anomalies and malfunctions