EasyManua.ls Logo

Getinge Volista - Installing and Removing the STG HLX Sterilisable Handle

Getinge Volista
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Use
Positioning the light
4
VOLISTA
IFU 01781 EN 19
57 / 106
Fig.56: Removing the STGPSX sterilisable
handle
Removing the sterilisable handle from a
lighthead
1. Press the locking button.
2. Remove the handle.
4.3.1.2 Installing and removing the STG HLX sterilisable handle
Fig.57: Installing the STGHLX sterilisable handle
Installing a sterilisable handle on the light-
head
1. Inspect the handle and check for cracks or
soiling.
2. Fit the handle to the mount.
3. Rotate the handle until its rotation is
locked.
Ø The locking button pops out of its hous-
ing.
Ø The handle is now locked in place and
ready for use.
Fig.58: Removing the STGHLX sterilisable
handle
Removing the sterilisable handle from a
lighthead
1. Press the locking button.
2. Remove the handle.