4
Use
Battery backup
84 / 106
VOLISTA
IFU 01781 EN 19
4.7.2 Performing battery tests
WARNING!
Risk of injury
A battery lifetime test fully discharges the batteries.
Do not perform an operation immediately after a battery lifetime test. Allow
time for the batteries to charge.
4.7.2.1 From the wall-mounted control keypad (on VCSII only)
12
13
14
Fig.101: Battery tests from the wall-mounted keypad
Running a battery backup test
1. Turn off the light.
2. Press Switchover test 12.
Ø If the test is successful, the battery level indicator 13 flashes green. If the test fails, the
battery level indicator 13 flashes red.
3. If the test fails, contact the Getinge technical service department.
4. Press Switchover test 12 again and hold until the button turns off.
Ø The light remains on at Level 3 and the system is ready for use.
Running a battery life test (only with a Getinge backup)
1. Turn off the light.
2. Press Battery lifetime test 14 and hold until the button is backlit.
Ø If the test is successful, the battery level indicator 13 flashes green. If the test fails, the
battery level indicator 13 flashes red.
3. If the test fails, contact the Getinge technical service department.
Ø The light turns off when the test is complete.
4. Press Battery lifetime test 14 again and hold until the button turns off.
NOTE
The battery life test can be stopped at any time by pressing Battery life test 14.