EasyManua.ls Logo

Giga-tronics 8003 - Two-Sensor Insertion Loss Measurements; Figure 2-31: Two Sensor Insertion Loss Setup W; a Power Splitter; Figure 2-32: Two Sensor Insertion Loss Using a Directional Coupler

Giga-tronics 8003
258 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
Manual 20791, Rev. C, June 2001 2-77
2.9.2 Two-Sensor Insertion Loss Measurements
The basic two-sensor insertion loss measurement consists of monitoring the power incident upon the
DUT, and comparing it to the power exiting the DUT. Incident power can be monitored in a variety of
ways, but the two most common are to use a Power Splitter or Directional Coupler, shown in
Figures 2-31 and 2-32.
Power splitters have the advantage of wide frequency range, typically from dc to tens of Gigahertz.
However, power splitters typically have an insertion loss of at least 6 dB. Directional couplers with an
insertion loss typically less than 1 dB are used when insertion loss is a concern. However, wide frequency
coverage is often more difficult and expensive than with power splitters.
Figure 2-31: Two Sensor Insertion Loss Setup w/ a Power Splitter
Figure 2-32: Two Sensor Insertion Loss using a Directional Coupler

Table of Contents

Related product manuals