Giga-tronics 8003 - Signal Sources; System Integrated; Operator Integrated; Modulation

Giga-tronics 8003
258 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction
Manual 20791, Rev. C, June 2001 1-15
1.6 Signal Sources
1.6.1 System Integrated
The 8003 can be system integrated (sweeper control using the 8003)
with the following sweepers:
Giga-tronics 12720A
Giga-tronics 7200 Series or 910 Series Synthesized Sweep
Generators
Giga-tronics 610 Series Sweep Oscillator
HP8350A/B with an RF plug-in (HP86200 series with the
HP11869A adapter)
HP8340A/B, HP8341A/B or HP8360 Synthesized Sweepers
Marconi 6300 Series Sweepers
Wiltron 6600B Sweep Generators
Wiltron 6700B Synthesized Sweep Generators
1.6.2 Operator Integrated
The 8003 is compatible with any signal source that meets the following requirements:
Horizontal Ramp: Provides a 0 to +10 V nominal ramp signal.
Blanking Signal: (Optional) Provides a TTL level during retrace and bandswitching.
1.6.3 Modulation
AC Detection Mode: A square wave is provided by the analyzer to modulate the signal
source.
Frequency: 1 kHz, nominal
On/Off Ratio: >30 dB
1.7 General Specifications
Temperature Range
Operating: 0°C to 50°C (32° to 122°F)
Storage: -40°C to 70°C (-40° to 158°F)
Power Requirements
100/120/220/240 Vac ±10%, 48 to 440 Hz, 200 VA typical
Physical Characteristics
Dimensions: 178 mm H x 451 mm W x 483 mm D (7.00 x 17.76 x 19.00 in)
Weight: 16.6 kg (36.5 lbs)