Introduction
Manual 20791, Rev. C, June 2001 1-15
1.6 Signal Sources
1.6.1 System Integrated
The 8003 can be system integrated (sweeper control using the 8003)
with the following sweepers:
• Giga-tronics 12720A
• Giga-tronics 7200 Series or 910 Series Synthesized Sweep
Generators
• Giga-tronics 610 Series Sweep Oscillator
• HP8350A/B with an RF plug-in (HP86200 series with the
HP11869A adapter)
• HP8340A/B, HP8341A/B or HP8360 Synthesized Sweepers
• Marconi 6300 Series Sweepers
• Wiltron 6600B Sweep Generators
• Wiltron 6700B Synthesized Sweep Generators
1.6.2 Operator Integrated
The 8003 is compatible with any signal source that meets the following requirements:
Horizontal Ramp: Provides a 0 to +10 V nominal ramp signal.
Blanking Signal: (Optional) Provides a TTL level during retrace and bandswitching.
1.6.3 Modulation
AC Detection Mode: A square wave is provided by the analyzer to modulate the signal
source.
Frequency: 1 kHz, nominal
On/Off Ratio: >30 dB
1.7 General Specifications
Temperature Range
Operating: 0°C to 50°C (32° to 122°F)
Storage: -40°C to 70°C (-40° to 158°F)
Power Requirements
100/120/220/240 Vac ±10%, 48 to 440 Hz, 200 VA typical
Physical Characteristics
Dimensions: 178 mm H x 451 mm W x 483 mm D (7.00 x 17.76 x 19.00 in)
Weight: 16.6 kg (36.5 lbs)