EasyManua.ls Logo

Giga-tronics 8003 - Key Operation; Figure 2-13: Softkey Menu Change from Frequency Change Prompt; START Key; UPDATE SWEEPER ON;OFF Key

Giga-tronics 8003
258 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8003 Precision Scalar Analyzer
2-52 Manual 20791, Rev. C, June 2001
2.6.4 Key Operation
2.6.4.1 UPDATE SWEEPER ON/OFF Key
[
UPD SWP ON/OFF
]
The UPD SWP ON/OFF (Update Sweeper On/Off) softkey function appears in all of the Source
Control menus. Its purpose is to enable or disable the source sweeper configuration update. The default
is ON (enable). For example: If it were desired to display a range of 1 to 10 GHz on the 8003 but the
sweeper output is from 2 to 5 GHz, the following procedure would be accomplished:
Set 2 to 5 GHz using the source Start and Stop functions
Turn OFF the UPD SWP function
Set the 8003 frequency for a range of 1 to 10 GHz
2.6.4.2 START Key
[START]
The SOURCE
[START]
key initiates source frequency changes by specifying the desired start, center,
or CW frequencies depending upon the mode of operation.
When
[START]
is pressed, the entry area of the display will show the current start, center, or CW
frequency depending upon whether the source is in the start/stop, center/
F, or CW mode, respectively.
A prompt will be displayed to change the frequency if desired, and the softkey menu will change to the
one illustrated in Figure 2-13.
[
START
], [
CENTER
], and [
CW
]
These softkeys change the mode and frequency of the source. The softkey label that corresponds to
the mode currently in use (START for Start/Stop, CENTER for Center/
F and CW) will be in
light blue. The other two options will be shown in standard menu green.
Pressing a softkey with a standard menu green label will cause the source to immediately switch to
measuring in the mode indicated by the softkey. The current value for the frequency associated with
NOTE:
A potential problem may occur when the 8003 and sweeper have different
configurations after the above is performed. Be sure to preset the instrument for
consistent operation.
Figure 2-13: Softkey Menu Change from Frequency Change Prompt
START
START
KEY
START
CENTER
CW
UPD SWP
ON/OFF

Table of Contents

Related product manuals