Remote Operation
Manual 20791, Rev. C, June 2001 3-19
Function Default Setting
Source
Signal Source Mode: Start/Stop Swept Freq
Start Frequency: Sweeper Start Frequency or 1 MHz if not
controlled by the 8003
Stop Frequency: Sweeper Stop Frequency
Sweep Time: 200 ms
RF Power Level: OFF, minimum RF level of the signal source or
0 dBm if not controlled by the 8003.
Source Modulation: OFF
The following instrument settings are not changed during an instrument preset:
All GPIB addresses (Analyzer, Source, Printers, and Plotters)
Sensor linearity calibrations
•Trace Memories
Path Cal Memories
3.3.10 Disable Private Bus Command Structure
DISABLE
Syntax: DISABLE
Related Command:
ENABLE
Description: This verb disables the private bus. When sent, the 8003 private bus becomes inactive. An
Enable command must be sent or a power down/power up cycle must be executed to
enable the private bus.
3.3.11 Enable Private Bus Command Structure
ENABLE
Syntax: ENABLE
Related Command:
DISABLE
Description: This verb enables the private bus. When sent, the 8003 private bus becomes active. A
Disable command must be sent to disable the private bus.