EasyManua.ls Logo

Giga-tronics 8003 - Sensor Path Calibration Command Structure

Giga-tronics 8003
258 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Operation
Manual 20791, Rev. C, June 2001 3-69
3.3.30 Sensor Path Calibration Command Structure
THRU, PATH
Syntax: THRU hs Sensor Specifier [us modifier] EOL
Example: PATH A;srq EOL
Specifiers: A, B, or C
Modifiers: This modifier is to request that a service request be sent upon completion of the Sensor
Path Calibration.
Modifiers name: Full Band, Narrow Band
Modifier Syntax 1: FB, NB
Modifier name: SRQ
Modifier Syntax 2: SRQ
If the SRQ modifier is received, the 8003 will path calibrate the specified sensor. When the
calibration is complete, the event complete bit will be set in the external status register and
the service request and the external status bits will be set in the instrument status byte.
Description: This verb defines and initiates a Sensor Path Calibration.

Table of Contents