Operation
Manual 20791, Rev. C, June 2001 2-105
After the above parameters have been entered, a display similar to Figure 2-59 will appear, showing the
selected settings and characteristics of the power sweep within the defined parameters.
A digital power sweep measurement can be made using the following keystrokes:
[POWER]
[POWER SWEEP] [PWRSWPON/OFF] (toggle ON)
[DIGITAL] (toggle to DIGITAL if necessary)
[STEP SIZE] (enter the desired step size in dBm)
[START POWER] (enter the desired starting power level in dBm)
[STOP POWER] (enter the desired stopping power level in dBm)
[FREQ] (enter the desired frequency)
If any other measurement modes are accessed, it may be necessary to reenter the Power Sweep Mode as
described above.
All of the usual measurements and corrections are possible (Ratio, Cal Factor, Averaging, Smoothing,
Frequency Correction, etc.)
Since the dynamic range of Giga-tronics power sensors is from -70 to +47 dBm depending on the sensor
type, the selected power range must be kept within these limits. It is up to you to make sure this criteria
is met.