EasyManua.ls Logo

Giga-tronics 8003 - Page 148

Giga-tronics 8003
258 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8003 Precision Scalar Analyzer
3-6 Manual 20791, Rev. C, June 2001
Typical HP-Basic Program for Stepped CW Measurements
100 ! ********************************************************
110 ! 8003 CW Measurement Demonstration Program
120 ! HP300 Series Version
130 ! Copyright 2001, Giga-tronics Power Measurements Division
140 ! *********************************************************
150 !
160 OPTION BASE 0! arrays in this example start with index of 0
170 DIM Buffer(19)! allocate storage for CW data
180 !
190 Status=SPOLL(704)! use serial poll to clear any existing srq s
200 CLEAR 704! Device Clear does a PRESET on the 8003
210 !
220 ! Set up analyzer channel configuration first
230 !
240 OUTPUT 704;POWER 1,A;avg 4;on! channel 1 is in CW mode, sensor A
250 ! and has an averaging number of 4
260 OUTPUT 704;CHAN 2;off;CHAN 3;off;CHAN 4;off
270 ! turn off channels 2, 3, and 4
280 !
290 ! Now, set up the sweeper parameters
300 !
310 OUTPUT 704;FIXED;freq 2000;level -15
320 ! Set CW frequency of 2 GHz and output level of -15 dBm
330 OUTPUT 704;FIXED;level on! turn on RF output
340 WAIT 8! put in appropriate wait time for sweeper to set up
350 !
360 ! This next line sets up the display format but is not strictly
370 ! necessary for data collection over the bus
380 !
390 OUTPUT 704;READOUT! make sure that the analyzer is in readout
400 ! (large number display) mode
410 !
420 ! Zero sensor A before taking a reading
430 !
440 ENABLE INTR 7;2! enable interrupts from instrument srq s
450 ON INTR 7 GOTO Done zeroa! exit from current routine when zeroing
done
460 OUTPUT 704;ZERO A;srq! zero sensor A, pull srq when done
470 Loop1: GOTO Loop1! Wait until srq (or do other tasks)
480 Done zeroa: Status=SPOLL(704)! Clear srq with serial poll
490 !
500 ! Take a series of readings in CW mode
510 !
520 FOR Reading no=0 TO 19! take 20 readings
530 OUTPUT 704;FIXED;freq ;2000+Reading no/19*(8000-2000)
540 ! set sweeper cw frequency to evenly space readings from 2 to 8 GHz
550 WAIT 1! put in time to make sure sweeper and analyzer are settled
560 OUTPUT 704;POWER 1,avg off! restart averaging by first turning off
570 OUTPUT 704;POWER 1;avg on! then turn averaging on
580 OUTPUT 704;OUTPUT 1! ask for reading
590 ENTER 704;Buffer(Reading no)! put the reading in the data buffer
600 NEXT Reading no
610 PRINT Buffer(*)
620 !
630 END

Table of Contents

Related product manuals