Remote Operation
Manual 20791, Rev. C, June 2001 3-15
SEARCH MOD: MIN
FULL Modifier Syntax: SCH[?] us MIN
query value returned: freq, data EOL
freq: freq = D.ddd Where d is zero to 6 digits. The value
represents the frequency of the minimum power point in
the trace in units of MHZ.
data: data =
±
D Where D is zero to three digits to the right of
the decimal point and three digits to the left. The
assumed units are dBm or dB depending upon the
measurement mode.
SEARCH MOD: RIGHT, RT
FULL Modifier Syntax: SCH[?] us RT hs numerical value
numerical value range: -90.00 to +90.00
numerical resolution: 0.01
assumed units: dBm, dB
query value returned: freq, data EOL
freq: freq = D.ddd Where D is zero to 6 digits. The value
represents the frequency of the cursor in units of MHZ.
This value is returned here for confirmation.
data: data =
±
D Where D is zero to three digits to the right of
the decimal point and three digits to the left. The
assumed units are dBm or dB depending upon the
measurement mode.
SEARCH MOD: LEFT, LT
FULL Modifier Syntax: SCH[?] us LT hs numerical
value numerical value range: -90.00 to +90.00
numerical resolution: 0.01
assumed units: dBm, dB
query value returned: freq, data EOL
freq: freq = D.ddd Where D is zero to 6 digits. The value
represents the frequency of the cursor in units of MHZ.
This value is returned here for confirmation.
data: data =
±
D Where D is zero to three digits to the right of
the decimal point and three digits to the left. The
assumed units are dBm or dB depending upon the
measurement mode.